Installation Procedure of the MO Sensor

Created by Patrícia Siqueira, Modified on Fri, 5 Jan at 10:10 AM by Patrícia Siqueira

If you encounter difficulties or problems in the operation of the system, we suggest checking the possible causes and simple solutions provided below. If this information is not sufficient to address the difficulty, please contact Treetech's Customer Support team or your authorized representative.

1. Remove the plastic cover that protects the end of the rod during transportation.

2. Loosen the nut (1) of the fastening element (2).
3. Push the fastening element (2) towards the lower end of the rod, as illustrated below.
4. Using a 22-sized wrench, thread and tighten the fastening element (2) onto the transformer valve. The use of Teflon tape is not required, as sealing is accomplished by the equipment's sealing ring (o-ring).
5. Carefully open the valve and insert the rod, adjusting its position until the sensor tip is in contact with the oil flow.

6. Using a 22-sized wrench, tighten the nut (1) on the fastening element (2), preventing any improper movement of the rod.


Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article