To check if the MO probe is functioning properly, follow the tests outlined in the table below:

Sensor Test - MO Probe (Humidity)
MO probe
Multimeter
Result
Range
4
-

-0,530
5


6
+
+0,545


MO probe
Multimeter
Result
Range
4
-

-0,670
5
+

6

+0,710


MO probe
Multímetro
Result
Range
4
+

Aberto
5


6
-
Aberto

 The multimeter should be set to the diode measurement mode.